Ablation catheter entrapment by chordae tendineae in the mitral valve during ventricular tachycardia ablation

Ki Hun Kim, Kee Joon Choi, Sung Hwan Kim, Gi Byoung Nam, You Ho Kim

Research output: Contribution to journalArticlepeer-review

5 Scopus citations

Abstract

Ablation Catheter Entrapment by Chordae Tendineae. We report on the entrapment of an ablation catheter by chordae tendineae in the mitral valve during radiofrequency (RF) ablation of ventricular tachycardia. The entrapped tip had to be removed via open surgery. Great care must be taken when performing radiofrequency ablation around the mitral valve apparatus.

Original languageEnglish
Pages (from-to)218-220
Number of pages3
JournalJournal of Cardiovascular Electrophysiology
Volume23
Issue number2
DOIs
StatePublished - Feb 2012

Keywords

  • catheter ablation
  • complication
  • mitral valve
  • ventricular tachycardia