Accuracy and precision of proton density fat fraction measurement across field strengths and scan intervals: A phantom and human study

  • Hye Jin Kim
  • , Hyo Jung Cho
  • , Bohyun Kim
  • , Myung Won You
  • , Jei Hee Lee
  • , Jimi Huh
  • , Jai Keun Kim

    Research output: Contribution to journalArticlepeer-review

    39 Scopus citations

    Abstract

    Background: Complex-based chemical shift imaging-based magnetic resonance imaging (CSE-MRI) is emerging as a preferred method for noninvasively quantifying proton density fat fraction (PDFF), a promising quantitative imaging biomarker (QIB) for longitudinal hepatic steatosis measurement. Purpose: To determine linearity, bias, repeatability, and reproducibility of the PDFF measurement using CSE-MRI (CSE-PDFF) across scan intervals, MR field strengths, and readers in phantom and nonalcoholic fatty liver disease (NAFLD) patients. Study Type: Institutional Review Board (IRB)-approved prospective. Subjects: Fat-water phantom and 20 adult patients. Field Strength/Sequence: 1.5 T and 3.0 T MR systems and a commercially available CSE-MRI sequence (IDEAL-IQ). Assessment: Two independent readers measured CSE-PDFF of fat-water phantom and NAFLD patients across two field strengths and scan intervals (same-day and 2-week) each and in a combination of both. MR spectroscopy-based PDFF (MRS-PDFF) was used as the reference standard for phantom PDFF. Statistical Tests: Linearity and bias of measurement were evaluated by linear regression analysis and Bland–Altman plots, respectively. Repeatability and reproducibility were assessed by coefficient of variance and repeatability / reproducibility coefficients (RC). The intraclass correlation coefficient was used to validate intra- and interobserver agreements. Results: CSE-PDFF showed high linearity and small bias (–0.6–0.4 PDFF%) with 95% limits of agreement within ±2.9 PDFF% across field strengths, 2-week interscan period, and readers in the clinical scans. CSE-PDFF was highly repeatable and reproducible both in phantom and clinical scans, with the largest observed RC across field strengths and 2-week interscan period being 3 PDFF%. Data Conclusion: CSE-PDFF is a robust QIB with high linearity, small bias, and excellent repeatability/reproducibility. A change of more than 3 PDFF% across field strengths within 2 weeks of scan interval likely reflects a true change, which is well within the clinically acceptable range. Level of Evidence: 3. Technical Efficacy: Stage 2. J. Magn. Reson. Imaging 2019;50:305–314.

    Original languageEnglish
    Pages (from-to)305-314
    Number of pages10
    JournalJournal of Magnetic Resonance Imaging
    Volume50
    Issue number1
    DOIs
    StatePublished - Jul 2019

    Bibliographical note

    Publisher Copyright:
    © 2018 International Society for Magnetic Resonance in Medicine

    Keywords

    • MRI
    • magnetic resonance imaging
    • nonalcoholic fatty liver disease
    • proton density fat fraction

    Fingerprint

    Dive into the research topics of 'Accuracy and precision of proton density fat fraction measurement across field strengths and scan intervals: A phantom and human study'. Together they form a unique fingerprint.

    Cite this