TY - JOUR
T1 - Analysis of Radiation Doses and Dose Reduction Strategies During Cerebral Digital Subtraction Angiography
AU - Yi, Ho Jun
AU - Sung, Jae Hoon
AU - Lee, Dong Hoon
AU - Kim, Sang Wook
AU - Lee, Sang Won
N1 - Publisher Copyright:
© 2017 Elsevier Inc.
PY - 2017/4/1
Y1 - 2017/4/1
N2 - Objective Adverse effects of increased use of cerebral digital subtraction angiography (DSA) include radiation-induced skin reactions and increased risk of malignancy. This study aimed to identify a method for reducing radiation exposure during routine cerebral DSA. Methods A retrospective review of 138 consecutive adult patients who underwent DSA with a biplane angiography system (Artis Zee, Siemens, Germany) from September 2015 to February 2016 was performed. In January 2016, the dose parameter was reset by the manufacturer from 2.4 μGy to 1.2 μGy. Predose (group 1) and postdose parameter reduction (group 2) groups were established. Angiograms and procedure examination protocols were reviewed according to patient age, gender, and diagnosis and angiography techniques were reviewed on the basis of the following radiation dose parameters: fluoroscopy time, reference point air kerma (Ka,r; in mGy), and kerma-area product (PKA; in μGym2). Results The mean Ka,r values in groups 1 and 2 were 1841.5 mGy and 1274.8 mGy, respectively. The mean PKA values in groups 1 and 2 were 23212.5 μGym2 and 14854.0 μGym2, respectively. Ka,r and PKA values were significantly lower in group 2 compared with group 1 (P < 0.001). Among individual factors, young age was a determining factor for reduced fluoroscopy time (P < 0.001), Ka,r (P = 0.047), and PKA (P = 0.022). Conclusions Increased awareness of radiation risks, as well as the establishment of strategies to reduce radiation dose, led to lower radiation doses for DSA. The use of appropriate examinations and low-dose parameters in fluoroscopy contributed significantly to the radiation dose reductions.
AB - Objective Adverse effects of increased use of cerebral digital subtraction angiography (DSA) include radiation-induced skin reactions and increased risk of malignancy. This study aimed to identify a method for reducing radiation exposure during routine cerebral DSA. Methods A retrospective review of 138 consecutive adult patients who underwent DSA with a biplane angiography system (Artis Zee, Siemens, Germany) from September 2015 to February 2016 was performed. In January 2016, the dose parameter was reset by the manufacturer from 2.4 μGy to 1.2 μGy. Predose (group 1) and postdose parameter reduction (group 2) groups were established. Angiograms and procedure examination protocols were reviewed according to patient age, gender, and diagnosis and angiography techniques were reviewed on the basis of the following radiation dose parameters: fluoroscopy time, reference point air kerma (Ka,r; in mGy), and kerma-area product (PKA; in μGym2). Results The mean Ka,r values in groups 1 and 2 were 1841.5 mGy and 1274.8 mGy, respectively. The mean PKA values in groups 1 and 2 were 23212.5 μGym2 and 14854.0 μGym2, respectively. Ka,r and PKA values were significantly lower in group 2 compared with group 1 (P < 0.001). Among individual factors, young age was a determining factor for reduced fluoroscopy time (P < 0.001), Ka,r (P = 0.047), and PKA (P = 0.022). Conclusions Increased awareness of radiation risks, as well as the establishment of strategies to reduce radiation dose, led to lower radiation doses for DSA. The use of appropriate examinations and low-dose parameters in fluoroscopy contributed significantly to the radiation dose reductions.
KW - Aneurysm
KW - Cerebral angiography
KW - Digital subtraction angiography
KW - Fluoroscopy
KW - Radiation dosage
UR - http://www.scopus.com/inward/record.url?scp=85012298653&partnerID=8YFLogxK
U2 - 10.1016/j.wneu.2017.01.004
DO - 10.1016/j.wneu.2017.01.004
M3 - Article
C2 - 28089806
AN - SCOPUS:85012298653
SN - 1878-8750
VL - 100
SP - 216
EP - 223
JO - World Neurosurgery
JF - World Neurosurgery
ER -