Abstract
The analytical performance and clinical application of measuring insulin and connecting peptide (C-peptide) by point of care (POC) assay were evaluated. A POC assay system (SelexOn, Osang Healthcare Inc., Anyang-si, Korea) was evaluated for precision, linearity, limit of blank (LOB), and limit of detection (LOD). Method comparison was performed with the Cobas Elecsys insulin and C-peptide assay (Roche Diagnostics GmbH, Mannheim, Germany) using 215 and 201 patient specimens for insulin and C-peptide, respectively. For clinical application, insulin resistance indices were studied. Homeostasis model assessment (HOMA) 1 and 2, Quantitative insulin sensitivity check index (QUIKI), fasting insulin resistance index (FIRI), and other indices were evaluated. The coefficient of variation (CV) of imprecision for low, medium, and high concentrations was 10.8%1, 15.99%, and 12.05%, respectively, for insulin and 9.21%, 13.51%, and 13.77%, respectively, for C-peptide. The linearity was validated to 839.78 pmol/L for insulin and to 17.30 nmol/L for C-peptide. LOB and LOD were 8.05 and 9.72 pmol/L for insulin and 0.05 and 0.08 nmol/L for C-peptide, respectively. For the method comparison, the regression equation was y = 1.259x − 8.818 (r = 0.957) for insulin and y = 1.163x–0.088 (r = 0.985) for C-peptide. The ROC value and overall accuracy were as follows: HOMA2 (C-peptide), 0.809, 79.7%; TyG, 0.788, 73.6%; CPR, 0.775, 74.8%; HOMA1, 0.725, 70.3%; QUIKI, 0.720, 70.3%; FIRI, 0.715, 70.1%; McAuley, 0.658, 65.1%; HOMA2 (Insulin), 0.645, 64.7%; Raynaud, 0.611, 61.4%, respectively. The POC assay system for insulin and C-peptide provided reliable results through a rapid and simple test that could be applied to clinical settings.
| Original language | English |
|---|---|
| Pages (from-to) | 347-353 |
| Number of pages | 7 |
| Journal | Scandinavian Journal of Clinical and Laboratory Investigation |
| Volume | 79 |
| Issue number | 5 |
| DOIs | |
| State | Published - 2019 |
Bibliographical note
Publisher Copyright:© 2019, © 2019 The Author(s). Published by Informa UK Limited, trading as Taylor & Francis Group.
UN SDGs
This output contributes to the following UN Sustainable Development Goals (SDGs)
-
SDG 3 Good Health and Well-being
Keywords
- C-peptide
- Insulin
- diabetes mellitus
- insulin resistance
- point-of-care testing
Fingerprint
Dive into the research topics of 'Analytical evaluation and clinical application of insulin and C-peptide by a whole blood, lateral flow, point of care (POC) assay system'. Together they form a unique fingerprint.Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver