BANG-3® polymer gel dosimetry in Cyberknife

  • Young Nam Kang
  • , Jisun Jang
  • , Byung Ock Choi
  • , Hong Suk Jang
  • , Ji Young Jung
  • , Hun Joo Shin
  • , Jae Hyuk Seo
  • , Ihl Bohng Choi
  • , Dong Joon Lee
  • , Soo Il Kwon
  • Catholic Univ. of Korea Coll. Med.
  • Kyonggi University
  • The Catholic University of Korea Incheon St. Mary's Hospital
  • Chungdam Wooridul Spine Hospital
  • Inje University

Research output: Chapter in Book/Report/Conference proceeding › Conference contribution › peer-review

Abstract

In this study, we performed BANG-3® polymer gel dosimetry to propose an experimental technique for relative output factor and 3D dose distribution measurements of small radiosurgical fields.We were irradiated 2 collimators (5 and 30 mm diameter) with 6 MV radiosurgery beam. And, MR scanned with the same slice thickness and three different in plane resolutions. In these experiments, output factor measurements with the Pin Point ion chamber and diode, EBT films are compared with BANG-3® polymer gel. We used a spherical phantom filled with BANG-3® polymer gel to measure 3D dose distributions. The irradiated phantom was scanned with an MRI scanner. Dose distributions were obtained by calibrating the polymer gel for a relationship between the absorbed dose and the spin-spin relaxation rate of the magnetic resistance signal. As results, the output factors founded agreement within ±1% as a reference the collimators 30 mm. Therefore, results suggest that polymer gel dosimetry coupled with the proposed experimental procedure helps overcome not only tissue-equivalence and beam perturbation implications but also volume averaging and positioning uncertainties in small radiosurgical beam dosimetry. In 3D dose distribution, the MRI artifacts cause a geometrical distortion of less than 2 mm and less than 10% change in the dose distribution. In conclusion, the gel dosimetry has the potential to qualitatively characterize the relative output factor and the dose distributions.

Original languageEnglish
Title of host publicationWorld Congress on Medical Physics and Biomedical Engineering
Subtitle of host publicationRadiation Oncology
PublisherSpringer Verlag
Pages683-685
Number of pages3
Edition1
ISBN (Print)9783642034725
DOIs
StatePublished - 2009
EventWorld Congress on Medical Physics and Biomedical Engineering: Radiation Oncology - Munich, Germany
Duration: 7 Sep 2009 → 12 Sep 2009

Publication series

NameIFMBE Proceedings
Number1
Volume25
ISSN (Print)1680-0737

Conference

ConferenceWorld Congress on Medical Physics and Biomedical Engineering: Radiation Oncology
Country/TerritoryGermany
CityMunich
Period7/09/09 → 12/09/09

Keywords

  • 3D dose distribution
  • BANG-3® polymer gel
  • CyberKnife
  • Output factor
  • Radisurgery