TY - JOUR
T1 - Comparison of the Abbott Architect i2000 assay, the Roche Modular Analytics E170 assay, and an immunoradiometric assay for serum hepatitis B virus markers
AU - Kim, Hyunjung
AU - Oh, Eun Jee
AU - Kang, Mi Sook
AU - Sung, Hoon Kim
AU - Park, Yeon Joon
PY - 2007/6
Y1 - 2007/6
N2 - Serum hepatitis B virus (HBV) markers are the most important data for epidemiological screening and clinical diagnosis of HBV infection, especially in endemic areas. We compared the results of the Roche Modular Analytics E170 assay, the Abbott Architect i2000 assay, and an immunoradiometric assay (IRMA) for HBV surface antigen (HBsAg), anti-HBV surface antigen (anti-HBs), HBV e antigen (HBeAg), and anti-HBV e antigen (anti-HBe). A number of serum samples (264, 263, 224, and 202 for HBsAg, anti-HBs, HBeAg, and anti-HBe, respectively) were studied. For samples giving discrepant results for HBeAg between methods, real-time PCR assays were performed. The concordance rates among the three methods were high for HBsAg (100%) and HBeAg (94.6), but low for anti-HBs (91.6%) and anti-HBe (82.2%). For anti-HBs, which could be measured quantitatively by the Modular E170 and Architect i2000 procedures, discrepant results were observed at low levels of anti-HBs. For anti-HBe, the positive rate was highest with Modular E170 (60.9%) followed by the IRMA kit (54.1%) and Architect i2000 (51.0%). This study shows substantial differences between the assay results by the three methods, which should be taken into account in determinations of serum HBV markers.
AB - Serum hepatitis B virus (HBV) markers are the most important data for epidemiological screening and clinical diagnosis of HBV infection, especially in endemic areas. We compared the results of the Roche Modular Analytics E170 assay, the Abbott Architect i2000 assay, and an immunoradiometric assay (IRMA) for HBV surface antigen (HBsAg), anti-HBV surface antigen (anti-HBs), HBV e antigen (HBeAg), and anti-HBV e antigen (anti-HBe). A number of serum samples (264, 263, 224, and 202 for HBsAg, anti-HBs, HBeAg, and anti-HBe, respectively) were studied. For samples giving discrepant results for HBeAg between methods, real-time PCR assays were performed. The concordance rates among the three methods were high for HBsAg (100%) and HBeAg (94.6), but low for anti-HBs (91.6%) and anti-HBe (82.2%). For anti-HBs, which could be measured quantitatively by the Modular E170 and Architect i2000 procedures, discrepant results were observed at low levels of anti-HBs. For anti-HBe, the positive rate was highest with Modular E170 (60.9%) followed by the IRMA kit (54.1%) and Architect i2000 (51.0%). This study shows substantial differences between the assay results by the three methods, which should be taken into account in determinations of serum HBV markers.
KW - Automated analyzers
KW - Hepatitis B viral markers
KW - Immunoradiometric assay
UR - https://www.scopus.com/pages/publications/34548146753
M3 - Article
C2 - 17709690
AN - SCOPUS:34548146753
SN - 0091-7370
VL - 37
SP - 256
EP - 259
JO - Annals of Clinical and Laboratory Science
JF - Annals of Clinical and Laboratory Science
IS - 3
ER -