Complete fracture of sirolimus-eluting stent in a saphenous vein graft to left anterior descending artery

Hong Yoo Sun, Won Jin Seung, Ho Her Sung, Jeoung Yoon Hee, Doo Kim Hyoung, Sun Im Yun, Bae Seung Ki, Hyung Kim Jae

Research output: Contribution to journalArticlepeer-review

1 Scopus citations

Abstract

Coronary stent fractures have been suggested as a potential new mechanism of restenosis. The mechanical properties of stents were designed not only to prevent vessel recoil, but also to resist the mechanical stress of vessel movement over millions of cardiac cycles. We present a case in which mechanical stress may have contributed to the fracture of a stent implanted in a saphenous vein graft (SVG) to the left coronary artery. The patient was admitted due to chest pain 2 years after receiving a coronary artery bypass graft. A coronary angiography revealed the culprit vessel to be the SVG to the left coronary artery. The graft was stenosed and was stented with a sirolimus-eluting stent. A 6-month follow-up coronary angiography revealed 80% in-stent restenosis with stent fracture. We re-intervened by balloon angioplasty. This is the first report of sirolimus-eluting stent fracture combined with restenosis of SVG in Korea.

Original languageEnglish
Pages (from-to)251-253
Number of pages3
JournalKorean Circulation Journal
Volume39
Issue number6
DOIs
StatePublished - Jun 2009

Keywords

  • Coronary artery bypass grafting
  • Drug-eluting stents

Fingerprint

Dive into the research topics of 'Complete fracture of sirolimus-eluting stent in a saphenous vein graft to left anterior descending artery'. Together they form a unique fingerprint.

Cite this