Effect of beam-flow angle on velocity measurements in modern doppler ultrasound systems

Michael Yong Park, Seung Eun Jung, Jae Young Byun, June Hoan Kim, Ga Eul Joo

Research output: Contribution to journalArticlepeer-review

47 Scopus citations

Abstract

OBJECTIVE. The purpose of this article is to examine the effect of different beam-flow angles on the accuracy of Doppler ultrasound velocity measurements in modern ultrasound systems. MATERIALS AND METHODS. A flow phantom was used to create a steady flow of water in a 4.3-mm-diameter tube. Using three different modern university-grade ultrasound systems, flow was measured at 30°, 40°, 50°, 60°, 70°, 80°, and 88° beam-flow angles twice by two radiologists in consensus using a convex and linear probe. Measured flow ratio, defined as measured velocity divided by estimated actual velocity, was calculated. Intraprobe, interprobe, and intermachine mean variation of measured flow ratio were calculated. RESULTS. Measured flow ratio increased as beam-flow angles increased. Measured flow ratios for the angles 30°, 40°, 50°, 60°, 70°, 80°, and 88° were 0.90, 0.97, 1.10, 1.22, 1.62, 2.34, and 10.29, respectively. Intraprobe, interprobe, and intermachine variation did not show marked differences. For angles grouped as 30-40°, 50-60°, 70°, and 80-88°, intraprobe variation was 12%, 15%, 15%, and 26%; interprobe variation was 20%, 16%, 13%, and 26%; and intermachine variation was 16%, 16%, 17%, and 54%, respectively. As beam-flow angle increased, an increase in spectral broadening was also noted. CONCLUSION. There is no simple cutoff beam-flow value, such as the well-quoted less than 60°, at which velocity measurements can be considered accurate. For follow-up imaging, beam-flow angle differences should be considered, and the same beam-flow angles should be used when possible. Follow-up imaging by different sonography machines is feasible.

Original languageEnglish
Pages (from-to)1139-1143
Number of pages5
JournalAmerican Journal of Roentgenology
Volume198
Issue number5
DOIs
StatePublished - May 2012

Keywords

  • Angle correction
  • Blood flow velocity
  • Doppler
  • Ultrasound

Fingerprint

Dive into the research topics of 'Effect of beam-flow angle on velocity measurements in modern doppler ultrasound systems'. Together they form a unique fingerprint.

Cite this