Effects of static dosimetric leaf gap on MLC-based small-beam dose distribution for intensity-modulated radiosurgery

  • Jeong Woo Lee
  • , Kyoung Sik Choi
  • , Semie Hong
  • , Yon Lae Kim
  • , Jin Beom Chung
  • , Doo Hyun Lee
  • , Bo Young Choe
  • , Hong Seok Jang
  • , Tae Suk Suh
  • Konkuk University
  • The Catholic University of Korea
  • The Catholic University of Korea, College of Medicine

Research output: Contribution to journalArticlepeer-review

24 Scopus citations

Abstract

The aim of the present study was to evaluate the effect of various specific dosimetric leaf gaps on the multileaf collimator (MLC)-based small-beam dose distribution. The dosimetric static leaf gap was determined by comparing the profiles of small MLC-based beams with those of small collimated fields (square fields of 1, 2, 3, and 4cm). The results showed that an approximately 2-mm gap was optimal with the Millennium 120-leaf MLC (Varian Medical Systems, Palo Alto, CA) and a Varian 21EX 6-MV photon beam. We also investigated how much the leaf gap affects the planning results and the actual dose distribution. A doughnut-shaped planning target volume (PTV, 6.1cm3) and inner organ at risk (OAR, 0.3cm3) were delineated for delicate intensity-modulated radiosurgery test planning. The applied leaf gaps were 0, 1, and 2 mm. The measured dose distributions were compared with the dose distribution in the treatment planning system. The maximum dose differences at inside PTV, outside PTV, and inner OAR were, respectively, 22.3%, 20.2%, and 35.2% for the 0-mm leaf gap; 17.8%, 22.8%, and 30.8% for the 1-mm leaf gap; and 5.5%, 8.5%, and 6.3 % for the 2-mm leaf gap. In a human head phantom (model 605: CIRS, Norfolk, VA) study, large dose differences of 1.3% - 12.7% were noted for the measurements made using the MLC files generated by the three different leaf gaps. The planned results were similar, and measurements showed a large dose difference associated with the various leaf gaps. These results strongly suggest that plans generated by a commercial inverse planning system commissioned using general collimated field data will probably demonstrate discrepancies between the planned treatments and the measured results.

Original languageEnglish
Pages (from-to)54-64
Number of pages11
JournalJournal of Applied Clinical Medical Physics
Volume8
Issue number4
DOIs
StatePublished - 2007

Keywords

  • Dosimetric static leaf gaps
  • MLC-based small beam

Fingerprint

Dive into the research topics of 'Effects of static dosimetric leaf gap on MLC-based small-beam dose distribution for intensity-modulated radiosurgery'. Together they form a unique fingerprint.

Cite this