Skip to main navigation Skip to search Skip to main content

Electrokinetic concentration on a microfluidic chip using polyelectrolytic gel plugs for small molecule immunoassay

  • Donghoon Han
  • , Kwang Bok Kim
  • , Yang Rae Kim
  • , Solji Kim
  • , Hee Chan Kim
  • , Junghwa Lee
  • , Joohoon Kim
  • , Taek Dong Chung
  • Seoul National University
  • Kyung Hee University

Research output: Contribution to journalArticlepeer-review

12 Scopus citations

Abstract

We have devised a new method to enhance the sensitivity of competitive immunoassay using electrokinetic concentration near a pair of highly charge-selective polymer [poly-2-acrylamido-2-methyl-1-propanesulfonic acid (pAMPSA)] plugs on a microfluidic chip. The polyelectrolytic gel, which was photopolymerized in a microfluidic channel network, served as the effective charge-selective extractor to sophisticatedly control the ion distribution. In this system, the fluorescent indicators on magnetic microbeads dispersed in the sample were spontaneously displaced by unlabeled target molecules and then electrokinetically preconcentrated in a single spot on the microfluidic chip. The locally preconcentrated fluorescent indicators were detected by laser-induced fluorescence. As a proof-of-concept, the competitive displacement assay of unlabeled 1 nM biotin was conducted to observe ca. 2000-fold enrichment within 3 min. In addition to the sensitive assessment of the unlabeled small target molecules, the proposed immunoassay system also showed good selectivity for biotin analogs such as biocytin, 2-iminobiotin, desthiobiotin.

Original languageEnglish
Pages (from-to)164-171
Number of pages8
JournalElectrochimica Acta
Volume110
DOIs
StatePublished - 2013

Bibliographical note

Funding Information:
This work was supported by the Public Welfare & Safety Research program (2010-0020772), Nano, Material Technology Development Program (2011-0030266), and the National Research Foundation of Korea (NRF) grant (No. 2012011289 ) funded by the Ministry of Education, Science and Technology (MEST) of Korea . J. Kim acknowledges the financial support from Basic Science Research Program through NRF funded by MEST of Korea (No. 2012R1A1A1014408 ).