Establishment of Local Diagnostic Reference Levels of Pediatric Abdominopelvic and Chest CT Examinations Based on the Body Weight and Size in Korea

Jae Yeon Hwang, Young Hun Choi, Hee Mang Yoon, Young Jin Ryu, Hyun Joo Shin, Hyun Gi Kim, So Mi Lee, Sun Kyung You, Ji Eun Park

Research output: Contribution to journalArticlepeer-review

13 Scopus citations

Abstract

Objective: The purposes of this study were to analyze the radiation doses for pediatric abdominopelvic and chest CT examinations from university hospitals in Korea and to establish the local diagnostic reference levels (DRLs) based on the body weight and size. Materials and Methods: At seven university hospitals in Korea, 2494 CT examinations of patients aged 15 years or younger (1625 abdominopelvic and 869 chest CT examinations) between January and December 2017 were analyzed in this study. CT scans were transferred to commercial automated dose management software for the analysis after being de-identified. DRLs were calculated after grouping the patients according to the body weight and effective diameter. DRLs were set at the 75th percentile of the distribution of each institution’s typical values. Results: For body weights of 5, 15, 30, 50, and 80 kg, DRLs (volume CT dose index [CTDIvol]) were 1.4, 2.2, 2.7, 4.0, and 4.7 mGy, respectively, for abdominopelvic CT and 1.2, 1.5, 2.3, 3.7, and 5.8 mGy, respectively, for chest CT. For effective diameters of < 13 cm, 14–16 cm, 17–20 cm, 21–24 cm, and >24 cm, DRLs (size-specific dose estimates [SSDE]) were 4.1, 5.0, 5.7, 7.1, and 7.2 mGy, respectively, for abdominopelvic CT and 2.8, 4.6, 4.3, 5.3, and 7.5 mGy, respectively, for chest CT. SSDE was greater than CTDIvol in all age groups. Overall, the local DRL was lower than DRLs in previously conducted dose surveys and other countries. Conclusion: Our study set local DRLs in pediatric abdominopelvic and chest CT examinations for the body weight and size. Further research involving more facilities and CT examinations is required to develop national DRLs and update the current DRLs.

Original languageEnglish
Pages (from-to)1172-1184
Number of pages13
JournalKorean Journal of Radiology
Volume22
Issue number7
DOIs
StatePublished - Jul 2021

Bibliographical note

Publisher Copyright:
© 2021 The Korean Society of Radiology.

Keywords

  • Child
  • Computed tomography
  • Diagnostic reference level
  • Radiation dose
  • Radiation protection

Fingerprint

Dive into the research topics of 'Establishment of Local Diagnostic Reference Levels of Pediatric Abdominopelvic and Chest CT Examinations Based on the Body Weight and Size in Korea'. Together they form a unique fingerprint.

Cite this