Abstract
Background: A novel enzymatic HbA1c assay was introduced for use in an automated chemistry analyzer. With this unique method, HbA1c and plasma glucose can be measured from the same EDTA tube. We evaluated the analytical performance of this enzymatic HbA1c assay in a JCA-BM6010/C analyzer and compared the HbA1c values with the results from other widely used methodological instruments. Methods: The imprecision, linearity, carry-over and concordance rate of the enzymatic HbA1c test (BM Test HbA1c) using the JCA-BM6010/C analyzer were evaluated. Three hundred and seventy-seven specimens with HbA1c concentrations from 16 to 133 mmol/mol were used for a comparison study with two high performance liquid chromatography methods: Variant II Turbo and Tosoh HLC 723 G8 and the AutoLab Hemoglobin A1c immunoturbidimetry reagent using a Hitachi 7600-110. Forty specimens were used for the glucose method comparison. Results: The HbA1c coefficients of variation of the within-run imprecision for low and high levels were 0.6% and 0.4%, respectively. The linearity of the BM Test HbA1c using the JCA-BM6010/C analyzer was excellent in the range between 31 mmol/mol and 143 mmol/mol. The carry-over rate was 0.2%. The relationships between the BM test and the other three methods were 0.916×Tosoh G8+3.644, r=0.986; 0.887×Bio-Rad Variant II+1.896, r=0.972; and 0.941×AutoLab+4.532, r=0.977. The concordance rates using a cut-off of 48 mmol/mol were 91.5% with Tosoh G8, 82.8% with Bio-Rad Variant II, and 91.0% with AutoLab. The simultaneously assayed plasma glucose with HbA 1c was 1.002×Routine plasma glucose+0.625, r=1.000 Conclusions: The enzymatic BM Test HbA1c in the JCA-BM6010/C analyzer showed excellent precision and linearity, and a minimal carry-over rate. The simultaneously assayed plasma glucose analysis showed good performance.
| Original language | English |
|---|---|
| Pages (from-to) | 2201-2208 |
| Number of pages | 8 |
| Journal | Clinical Chemistry and Laboratory Medicine |
| Volume | 51 |
| Issue number | 11 |
| DOIs | |
| State | Published - 1 Nov 2013 |
Bibliographical note
Funding Information:stated that there are no conflicts of interest regarding the publication of this article. Research funding played no role in the study design; in the collection, analysis, and interpretation of data; in the writing of the report; or in the decision to submit the report for publication. Research funding: This study was supported by a grant from Alere Healthcare Inc., Seoul, Korea. Employment or leadership: None declared. Honorarium: None declared.
UN SDGs
This output contributes to the following UN Sustainable Development Goals (SDGs)
-
SDG 3 Good Health and Well-being
Keywords
- HbA
- diabetes
- enzymatic HbA assay
- evaluation
- glucose
- multicenter study
Fingerprint
Dive into the research topics of 'Evaluation of enzymatic BM Test HbA1c on the JCA-BM6010/C and comparison with Bio-Rad Variant II Turbo, Tosoh HLC 723 G8, and AutoLab immunoturbidimetry assay'. Together they form a unique fingerprint.Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver