Identification of a frit-related sample carryover in newborn screening by tandem mass spectrometry

Kyunghun Kim, Howon Lee, Jeong Joong Lee, Kyoungho Cha, Nam Hyun Park, Young Keun Shin, Hyojin Chae, Eun Jee Oh

Research output: Contribution to journalArticlepeer-review

1 Scopus citations

Abstract

The need for high-throughput analysis of multiple analytes for inborn errors of metabolism in newborn screening (NBS) has led to the introduction of tandem mass spectrometry (MS/MS) into the NBS laboratory. In a flow-injection analysis (FIA), the predominant MS/MS method utilized for NBS, samples are introduced directly into the mass spectrometer without chromatographic separation. When a high-throughput FIA-based MS/MS method is implemented on newer generations of mass spectrometers with increased sensitivity, the risk of carryover and contamination increases. In the present study, we report the carryover of ornithine identified during the implementation of the NeoBase™ 2 (PerkinElmer) non-derivatized kits on the Xevo-TQD platform (Waters Corporation) and describe the source of the carryover, which was traced to the stainless-steel frit-type inline filter. Furthermore, a possible compound-dependent interaction with the stainless-steel frit is suggested based on the structure of ornithine and its effect on separation techniques. Investigation and mitigation of carryover can be a time and resource consuming process, and to this end, our report on identification of a stainless-steel frit as the source of delayed elution and carryover of ornithine should be recognized as a rare, albeit possible source of carryover in FIA-MS/MS methods adopted for NST.

Original languageEnglish
Pages (from-to)56-60
Number of pages5
JournalJournal of Mass Spectrometry and Advances in the Clinical Lab
Volume27
DOIs
StatePublished - Jan 2023

Bibliographical note

Funding Information:
None.

Publisher Copyright:
© 2023 THE AUTHORS

Keywords

  • Carryover
  • Flow injection analysis
  • Newborn screening
  • Ornithine
  • Tandem mass spectrometry

Fingerprint

Dive into the research topics of 'Identification of a frit-related sample carryover in newborn screening by tandem mass spectrometry'. Together they form a unique fingerprint.