Objective: This study aimed to evaluate the image quality and dose reduction of low-dose three-dimensional (3D) rotational angiography (RA) for evaluating intracranial aneurysms. Materials and Methods: We retrospectively evaluated the clinical data and 3D RA datasets obtained from 146 prospectively registered patients (male:female, 46:100; median age, 58 years; range, 19–81 years). The subjective image quality of 79 examinations obtained from a conventional method and 67 examinations obtained from a low-dose (5-seconds and 0.10-μGy/frame) method was assessed by two neurointerventionists using a 3-point scale for four evaluation criteria. The total image quality score was then obtained as the average of the four scores. The image quality scores were compared between the two methods using a noninferiority statistical testing, with a margin of-0.2 (i.e., score of low-dose group – score of conventional group). For the evaluation of dose reduction, dose-area product (DAP) and air kerma (AK) were analyzed and compared between the two groups. Results: The mean total image quality score ± standard deviation of the 3D RA was 2.97 ± 0.17 by reader 1 and 2.95 ± 0.20 by reader 2 for conventional group and 2.92 ± 0.30 and 2.95 ± 0.22, respectively, for low-dose group. The image quality of the 3D RA in the low-dose group was not inferior to that of the conventional group according to the total image quality score as well as individual scores for the four criteria in both readers. The mean DAP and AK per rotation were 5.87 Gy-cm2 and 0.56 Gy, respectively, in the conventional group, and 1.32 Gy-cm2 (p < 0.001) and 0.17 Gy (p < 0.001), respectively, in the low-dose group. Conclusion: Low-dose 3D RA was not inferior in image quality and reduced the radiation dose by 70%–77% compared to the conventional 3D RA in evaluating intracranial aneurysms.

Original languageEnglish
Pages (from-to)256-263
Number of pages8
JournalKorean Journal of Radiology
Volume23
Issue number2
DOIs
StatePublished - Feb 2022

Bibliographical note

Publisher Copyright:
© 2022 The Korean Society of Radiology.

Keywords

  • Image quality
  • Intracranial aneurysm
  • Low-dose 3D rotational angiography
  • Radiation dose

Fingerprint

Dive into the research topics of 'Low-Dose Three-Dimensional Rotational Angiography for Evaluating Intracranial Aneurysms: Analysis of Image Quality and Radiation Dose'. Together they form a unique fingerprint.

Cite this