Measurement of pupil centroid shift and cyclotorsional displacement using iris registration

  • Catholic Univ. of Korea Coll. Med.
  • Korean Eye and Gene Bank Related to Blindness

Research output: Contribution to journalArticlepeer-review

41 Scopus citations

Abstract

Background/Aims: To estimate the amount and nature of pupil centroid shift and cyclotorsion occurring between preoperative wavefront measurement and refractive surgery. Methods: This retrospective study comprised 183 eyes of 92 patients who had wavefront-guided LASIK using iris registration. We analyzed the pupil centroid shift and cyclotorsional displacement occurring between preoperative wavefront measurements and during the laser procedure. The eyes were measured by the WaveScan® Wavefront® system (VISX, Santa Clara, Calif., USA) and refractive surgery was performed with the VISX STAR S4 excimer laser system. Results: The mean magnitude of shifts in the location of pupil center was 0.26 ± 0.16 mm horizontally, 0.20 ± 0.16 mm vertically and 0.37 ± 0.13 mm vectorally. 162 eyes (88.5%) experienced more than 0.2 mm of shift. The mean amount of torsion was 2.58 ± 1.56° for 83 eyes showing incyclotorsion and 2.94 ± 1.87° for 100 eyes showing excyclotorsion. 117 eyes (63.9%) experienced more than 2° of cyclotorsion and 21 eyes (11.5%) experienced more than 5° of cyclotorsion. Conclusions: Pupil centroid shift and cyclotorsion occur in the majority of eyes undergoing laser treatment. The amount of pupil centroid shift and cyclotorsion shown in this study is considerable. Accurate compensation for cyclotorsion and pupil centroid shift may maximize the potential of visual quality of customized ablations.

Original languageEnglish
Pages (from-to)166-171
Number of pages6
JournalOphthalmologica
Volume223
Issue number3
DOIs
StatePublished - May 2009

Keywords

  • Centroid shift
  • Cyclotorsional displacement
  • Refractive surgery
  • Wavefront measurement

Fingerprint

Dive into the research topics of 'Measurement of pupil centroid shift and cyclotorsional displacement using iris registration'. Together they form a unique fingerprint.

Cite this