Abstract

Background: The utilization of short humeral stems in reverse total shoulder arthroplasty has gained attention in recent times. However, concerns regarding the risk of misalignment during implant insertion are associated with their use. Methods: Eight fresh-frozen cadaveric shoulders were prepared for dissection and biomechanical testing. A bespoke humeral implant was fabricated to facilitate assessment of neutral, varus, and valgus alignments using a single stem, and 10° was established as the maximum permissible angle for misalignments. Shift in humerus position and changes in deltoid length attributable to misalignments relative to the neutral position were evaluated using a Microscribe 3DLx system. The impingement-free range of motion, encompassing abduction, adduction, internal rotation, and external rotation (ER), was gauged using a digital goniometer. The capacity for abduction was evaluated at maximal abduction angles under successive loading on the middle deltoid. A specialized traction system coupled with a force transducer was employed to measure anterior dislocation forces. Results: Relative to the neutral alignment, valgus alignment resulted in a more distal (10.5 ± 2.4 mm) and medial (8.3 ± 2.2 mm) translation of the humeral component, whereas the varus alignment resulted in the humerus shifting more superiorly (11.2 ± 1.3 mm) and laterally (9.9 ± 0.9 mm) at 0° abduction. The valgus alignment exhibited the highest abduction angle than neutral alignment (86.2°, P < .001). Conversely, the varus alignment demonstrated significantly higher adduction (18.4 ± 7.4°, P < .001), internal rotation (68.9 ± 15.0°, P = .014), and ER (45.2 ± 10.5°, P = .002) at 0° abduction compared to the neutral alignments. Anterior dislocation forces were considerably lower (23.8 N) in the varus group compared to the neutral group at 0°ER (P = .047). Additionally, abduction capability was markedly higher in varus alignment at low deltoid loads than the neutral alignment (5N, P = .009; 7.5 N, P = .007). Conclusions: The varus position enhances rotational range of motion (ROM) but increases instability, while the valgus position does not significantly impact ROM or instability compared to the neutral position.

Original languageEnglish
Pages (from-to)1762-1770
Number of pages9
JournalJournal of Shoulder and Elbow Surgery
Volume33
Issue number8
DOIs
StatePublished - Aug 2024

Bibliographical note

Publisher Copyright:
© 2024

Keywords

  • Basic Science Study
  • Biomechanics
  • Short humeral stem
  • biomechanics
  • misalignment
  • misalignment of stem
  • reverse total shoulder arthroplasty

Fingerprint